REGENERATIVE BATTERY CYCLER

Bidders ID: NBD00159225297387987
Due Date: May 8, 2026
Posted Date: Jul 23, 2026
Level of Government: Federal
State: Indiana
Agency: DEPT OF DEFENSE
Category:
  • 61 - Electric Wire, and Power and Distribution Equipment
Solicitation Number: N0016426SNB96
Source: Members Only
Follow
REGENERATIVE BATTERY CYCLER
Active
Contract Opportunity
Notice ID
N0016426SNB96
Related Notice
Department/Ind. Agency
DEPT OF DEFENSE
Sub-tier
DEPT OF THE NAVY
Major Command
NAVSEA
Sub Command
NAVSEA WARFARE CENTER
Office
NSWC CRANE
General Information
  • Contract Opportunity Type: Special Notice (Original)
  • Original Published Date: Jul 23, 2026 04:42 pm EDT
  • Original Response Date: May 08, 2026 12:00 pm EDT
  • Inactive Policy: 15 days after response date
  • Original Inactive Date: May 23, 2026
  • Initiative:
    • None
Classification
  • Original Set Aside: No Set aside used
  • Product Service Code: 6130 - CONVERTERS, ELECTRICAL, NONROTATING
  • NAICS Code:
    • 334515 - Instrument Manufacturing for Measuring and Testing Electricity and Electrical Signals
  • Place of Performance:
    IN 47522
    USA
Description

MARKET SURVEY - SOURCES SOUGHT SYNOPSIS: The Government is issuing this RFl/sources sought announcement as part of market survey to identify qualified sources capable of providing a regenerative battery cycler to replace NHR 9200 cyclers. This equipment will be utilized to test, cycle, and evaluate large-scale lead-acid battery systems to include VRLA chemistries. used in mission-critical facility backup and maritime applications. To enable seamless regenerative discharge and recharge cycles, the equipment must function as a bi-directional DC source and load.





The Contractor shall provide a turnkey testing system, including all necessary power hardware, control software, environmental sensor integration, and safety interlocks required to safely cycle high-capacity lead-acid battery banks.



Interested sources must demonstrate the capability to provide a system that meets or exceeds the following specifications:





Electrical & Performance Specifications





Parameter



Required Specification



Voltage Range



1 - 40V



Current Range



0 - 3000A (Charge and Discharge)



Power Range (Charge)



0 - 40 kW



Power Range (Discharge)



0 - 60 kW



Facility Input Power



480 VAC, 3-Phase, 60 Hz, 80A



Cooling



Air Cooling is preferred



Regenerative Power Efficiency



Regenerative discharge capabilities of at least 90%





AC Ripple Limitation



AC ripple voltage imposed on the DC bus must be less than 0.5% RMS of the float voltage



Voltage Regulation Precision



Maintain a steady-state DC output voltage regulation of +/- 0.5% (or tighter) across all operating ranges, independent of line voltage fluctuations or rapid load transitions.





Design & Integration Capabilities





Feature



Requirement



System Architecture



Bi-directional DC source/load and battery cycling equipment utilizing a modular design.



Parallel Channel Scaling



Capable of paralleling multiple channels to multiply the current capacity



Control Modes



System must support Constant Voltage (CV), Constant Current (CC), and Constant Power (CP) along with pulsing operational modes.



Hardware Integration



Must be fully compatible with National Instruments (NI) CompactRIO (cRIO) controllers and data acquisition (DAQ) systems.



Software Integration



Must be LabVIEW compatible and support standard integration utilizing Interchangeable Virtual Instrument (IVI) drivers.































The system must incorporate multiple independent layers of hardware and software safety controls to protect personnel, facility infrastructure, and the Unit Under Test (UUT). The system must meet or exceed the following safety capabilities:





Safety Feature



Requirement



Physical Isolation



Must feature internal, independent electromechanical safety contactors on the DC output channels.



Hardware Interlocks



Must include a dedicated, dual-channel external wired interlock loop for integration with thermal chambers or external safety gates, alongside a front-mounted physical Emergency Stop (E-Stop) Interlocks.



UUT Protection Circuits



Must feature integrated reverse-polarity checking and a soft-start pre-charge circuit to prevent in-rush current surges upon battery connection.



Programmable Protection



Software and firmware must allow users to define strict, fast-acting trip thresholds for Over-Voltage (OV), Under-Voltage (UV), Over-Current (OC), and Over-Power (OP).



Communication Fail-Safe



Must feature a programmable hardware watchdog timer that automatically shuts down the outputs and opens isolation contactors if communications with the master controller (e.g., cRIO) are interrupted.





Warranty and Maintenance:





Support Category



Requirement



Standard Warranty



Provide details on the standard OEM warranty (minimum 12 months preferred), including coverage details (parts, labor, and travel) and return-to-factory vs. on-site repair terms.



Preventive Maintenance



Describe available preventive maintenance agreements (PMAs), including the recommended frequency of on-site inspections, fan/filter replacements, and thermal checks.



Calibration Services



Detail the requirements and procedures for system calibration. The system must be capable of being calibrated to ISO/IEC 17025:2017 standards. Specify if calibration can be performed on-site or if the unit must be returned to the manufacturer.



Technical Support



Provide information on available technical support structures, including guaranteed response times (e.g., 24/48-hour response), availability of remote diagnostics, and software update access.





The technical specifications listed represent the Government's target operational baseline. If interested sources cannot meet these exact thresholds (e.g., specific voltage, current, or power ranges), they are requested to submit their standard commercial capabilities, highlighting where their system deviates, and propose alternative configurations that could meet the overall mission objective of replacing the NHR 9200-60 with equivalent or upgraded equipment.







Responses should include the following:





Include your companies technical and manufacturing capabilities and quality certifications held. Lastly, please provide a history of your company and past performance on Government contracts if any.





This submission may be in the form of product literature, operating instructions, and/or supplemental technical documentation that define the item(s) and associated hardware. Potential sources should provide any delivery, shipping or warranty information.



Responses from Small and Small Disadvantaged Business firms are highly encouraged. Firms responding should indicate if they are a small business, a socially and economically disadvantaged business, 8(a) firms, historically black colleges or universities, and minority institutions.



Responses to the RFI shall include the following:



Submitter's Name



Street Address, City, State, NINE-Digit zip code



Point of Contact (POC)



Cage Code



Size and Socioeconomic Status



POC Telephone and facsimile numbers



Announcement Number: N0016426SNB96





Contractors must be properly registered in the Government System for Award Management database (SAM). Offerors may obtain information on SAM registration and annual confirmation requirements by calling 1-866-606-8220, or via the Internet at www.sam.gov.



At the current time, NO FUNDING IS AVAILABLE for contractual efforts. This sources sought synopsis does not constitute an Invitation for Bids (IFB), a Request for Quote (RFQ), or a Request for Proposal (RFP) and it should not be construed as a commitment of any kind by the government to issue a formal solicitation or ultimately award a contract. The US Government is in no way liable to pay for or reimburse any companies or entities that respond to this announcement. Any costs incurred by interested companies in response to this announcement will NOT be reimbursed.



Please refer to announcement number N0016426SNB96 in all correspondence and communications. E-mail point of contact for information regarding the subject RFI is Amy Haworth; amy.n.haworth.civ@us.navy.mil and Andrew Ficklin; andrew.b.ficklin.civ@us.navy.mil.



All changes that occur will be posted to the beta.SAM.gov website.


Attachments/Links
Contact Information
Contracting Office Address
  • WARFARE CENTER 300 HIGHWAY 361
  • CRANE , IN 47522-5001
  • USA
Primary Point of Contact
Secondary Point of Contact
History
  • Jul 23, 2026 04:42 pm EDTSpecial Notice (Original)
Disclaimer: The details regarding bids, requests for proposals (RFPs), and requests for qualifications (RFQs) on this site are provided for convenience and do not represent official public notice. To respond to or inquire about bids, RFPs, or RFQs, please contact the appropriate government department.

TRY FOR FREE

Not a GovernmentBidders Member Yet?

Get unlimited access to thousands of active local, state and federal government bids and awards in All 50 States.

Start Free Trial Today >